IMPLANT TORQUE · ZIMMER BIOMET

Zimmer Biomet torque values.

One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Zimmer Biomet documentation, with the document named beside it.

Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
CONNECTIONSTapered Screw-Vent internal hex (TSV · TM · TSX) · 3i Certain internal · 3i External hex
LINES COVEREDTapered Screw-Vent (TSV) · Trabecular Metal (TM) · TSX · 3i T3 / T3 PRO · Osseotite Certain · 3i External Hex
TrazaLabTORQUE SCALE Zimmer BiometManufacturer document
EDITORIAL VISUALIZATION
≤ 90Ncm
PLACEMENTTSX — instrument limits

“Do not exceed 90 Ncm when pulling on the torque-indicating arm.” Latchlock Gemlock drivers: do not exceed 80 Ncm.

DRIVERZTIRW torque-indicating wrench · Latchlock drivers ≤ 80 Ncm
ZimVie TSX Implant Surgical Manual (ZVINST0015, REV F 12/25)

THE VALUES

Every component, its number, its document.

Rows are grouped in treatment order. Where Zimmer Biomet publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.

ComponentApplies toTorqueFrom the documentSource
PlacementInsertion torque of the fixtureTSV · TM — recommended rangenot publishedThe current TSV/TM surgical manual publishes no recommended insertion torque or maximum; it only notes that cortical bone taps may be used to reduce insertion torque.ZimVie TM & TSV Surgical Manual (ZVINST0082, REV C 12/25)Not published — confirm IFU
PlacementInsertion torque of the fixtureTSX — instrument limits
Driver: ZTIRW torque-indicating wrench · Latchlock drivers ≤ 80 Ncm
≤ 90Ncm“Do not exceed 90 Ncm when pulling on the torque-indicating arm.” Latchlock Gemlock drivers: do not exceed 80 Ncm.ZimVie TSX Implant Surgical Manual (ZVINST0015, REV F 12/25)Manufacturer document
PlacementInsertion torque of the fixture3i T3 · Certain · External hex> 50 hand · > 90 tapNcmAbove 50 Ncm switch to hand ratcheting; above 90 Ncm tapping is required. Exceeding 90 Ncm may deform or strip the driver tip or the internal hex. (Biomet 3i era document.)BIOMET 3i Tapered & Parallel Walled Surgical Manual (INSTSM REV F 06/17)Manufacturer document
HealingHealing abutment · cover screwTSV · TM cover screw and healing collar
Driver: 1.25 mm hex
hand-tight“Tighten using finger pressure only.” No Ncm value is published.ZimVie TM & TSV Surgical Manual (ZVINST0082, REV C 12/25)Manufacturer document
HealingHealing abutment · cover screw3i Certain: cover screw 10 · healing abutment 20
Driver: Large Hex Driver (PHD02N)
10 · 20NcmEncode and standard healing abutments 20 Ncm; Low Profile and conical healing caps 10 Ncm.ZimVie T3 / T3 PRO / Osseotite Restorative Manual (ZVINST0102, REV D 12/25)Manufacturer document
AbutmentFinal abutment screw at implant levelTSV · TM · TSX, implant level, single unit
Driver: 1.25 mm hex · Restorative Torque Wrench (TWR, 10–35 Ncm)
30NcmEngaging single-unit Ti-Base and abutment screws: recommended torque 30 Ncm.ZimVie Prosthetic Overview TSX / TM / TSV / Eztetic (ZV0796, REV G 05/26)Manufacturer document
AbutmentFinal abutment screw at implant level3i Certain — Gold-Tite hexed screw
Driver: Large Hex Driver
20Ncm“Torque the Certain Gold-Tite Hexed Screw to 20 Ncm.”ZimVie T3 / T3 PRO / Osseotite Restorative Manual (ZVINST0102, REV D 12/25)Manufacturer document
AbutmentFinal abutment screw at implant level3i External hex — Gold-Tite square screw
Driver: Square Driver
32–35Ncm“Torque the Gold-Tite Square Screw to 32–35 Ncm.” Gold slotted screws: 10 Ncm.ZimVie T3 / T3 PRO / Osseotite Restorative Manual (ZVINST0102, REV D 12/25)Manufacturer document
Multi-unitMulti-unit abutment into the implantTSV Tapered Abutment → implant, straight and angled
Driver: Restorative Torque Wrench (TWR)
30Ncm“Tighten the angled tapered abutment retaining screws and the straight tapered abutments to 30 Ncm.” 3i Low Profile abutments: 20 Ncm straight and angled.ZimVie Revitalize Surgical & Restorative Technique Guide (ZVINST0152)Manufacturer document
Prosthetic screwScrew that holds the prosthesis on the abutmentTSV Tapered Abutment coping screws · abutment-level Ti-Base
Driver: Restorative Torque Wrench (TWR)
20Ncm“Torque the coping screws to 20 Ncm.” 3i Low Profile retaining screws: 10 Ncm.ZimVie Revitalize Surgical & Restorative Technique Guide (ZVINST0152)Manufacturer document
TemporaryTemporary abutment or copingTSV implant-level temporary abutment
Driver: 1.25 mm hex
not publishedNot published in the ZimVie documents we read. Tapered Abutment healing caps and transfer copings: hand-tight with the 1.25 mm hex driver.ZimVie Revitalize Surgical & Restorative Technique Guide (ZVINST0152)Not published — confirm IFU

Re-tightening, as the manufacturer states it

ZimVie’s Revitalize guide says to retighten the Tapered Abutments to 30 Ncm after the healing caps are removed for the impression — a procedural step, not a timed interval. No “after X minutes” statement was found.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

ONE SITE, FIVE MOMENTS

Where the numbers appear in the treatment.

Site 36, one fixture, one crown. The same Zimmer Biomet values from the table, placed on the moment each one is read off the wrench.

EDITORIAL VISUALIZATION
  1. 01PLACEMENT≤ 90NcmFixture seated

    TSX — instrument limits

  2. 02HEALING10 · 20NcmHealing abutment on

    3i Certain: cover screw 10 · healing abutment 20

  3. 03ABUTMENT30NcmFinal abutment torqued

    TSV · TM · TSX, implant level, single unit

  4. 04PROSTHESIS20NcmCrown screw seated

    TSV Tapered Abutment coping screws · abutment-level Ti-Base

  5. 05RE-TIGHTENING30NcmSame value, second pass

    TSV · TM · TSX, implant level, single unit

Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.

AFTER THE CLICK

The value you applied is the value the laboratory should read.

In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.

See the Digital Rx
TrazaLabTREATMENT DESIGN · TOOTH 36Digital Rx
TREATMENT DESIGN · TOOTH 36
IMPLANT SYSTEMZimmer Biomet
PLATFORMTapered Screw-Vent (TSV)
CONNECTIONTapered Screw-Vent internal hex (TSV
ABUTMENTFinal abutment
TORQUE APPLIED30 Ncm
DIGITAL RXTorque applied · 30 Ncm
LABORATORY · PER-TOOTH SHEET36 · implant crown · torque 30 Ncm

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01Is the Zimmer TSV abutment screw 30 or 35 Ncm?+

30 Ncm. ZimVie’s prosthetic overview and technique guides give 30 Ncm for the single-unit implant-level screw and for the Tapered Abutment. Third-party tables that list 35 Ncm for TSV contradict the manufacturer; the 3i Certain hexed screw is 20 Ncm.

02Why do 3i Certain and TSV use different torque values?+

They are different connections with different screws. The Certain internal connection restores with a Gold-Tite hexed screw at 20 Ncm; the Screw-Vent internal hex restores at 30 Ncm; the 3i external hex keeps its Gold-Tite square screw at 32–35 Ncm. Identify the connection on the case record before you set the wrench.

03Do these Zimmer Biomet values replace the IFU?+

No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.

04Should I re-tighten after ten minutes?+

The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.

05Where does the applied torque go after the appointment?+

In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.