One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Zimmer Biomet documentation, with the document named beside it.
ZimVie TSX Implant Surgical Manual (ZVINST0015, REV F 12/25)
THE VALUES
Every component, its number, its document.
Rows are grouped in treatment order. Where Zimmer Biomet publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.
Component
Applies to
Torque
From the document
Source
PlacementInsertion torque of the fixture
TSV · TM — recommended range
not published
The current TSV/TM surgical manual publishes no recommended insertion torque or maximum; it only notes that cortical bone taps may be used to reduce insertion torque.
Above 50 Ncm switch to hand ratcheting; above 90 Ncm tapping is required. Exceeding 90 Ncm may deform or strip the driver tip or the internal hex. (Biomet 3i era document.)
TSV Tapered Abutment → implant, straight and angled
Driver: Restorative Torque Wrench (TWR)
30Ncm
“Tighten the angled tapered abutment retaining screws and the straight tapered abutments to 30 Ncm.” 3i Low Profile abutments: 20 Ncm straight and angled.
ZimVie’s Revitalize guide says to retighten the Tapered Abutments to 30 Ncm after the healing caps are removed for the impression — a procedural step, not a timed interval. No “after X minutes” statement was found.
Newest revision read: 2026. Manufacturers revise these documents; the revision printed on your IFU wins.
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
WHY THE SCREW LOOSENS
You torqued it. Then the metal settled.
Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.
Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.
AFTER THE CLICK
The value you applied is the value the laboratory should read.
In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
01Is the Zimmer TSV abutment screw 30 or 35 Ncm?+
30 Ncm. ZimVie’s prosthetic overview and technique guides give 30 Ncm for the single-unit implant-level screw and for the Tapered Abutment. Third-party tables that list 35 Ncm for TSV contradict the manufacturer; the 3i Certain hexed screw is 20 Ncm.
02Why do 3i Certain and TSV use different torque values?+
They are different connections with different screws. The Certain internal connection restores with a Gold-Tite hexed screw at 20 Ncm; the Screw-Vent internal hex restores at 30 Ncm; the 3i external hex keeps its Gold-Tite square screw at 32–35 Ncm. Identify the connection on the case record before you set the wrench.
03Do these Zimmer Biomet values replace the IFU?+
No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.
04Should I re-tighten after ten minutes?+
The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.
05Where does the applied torque go after the appointment?+
In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.