  2. Implant torque

IMPLANT TORQUE BY BRAND

Nine implant systems. Every torque, sourced.

Placement, healing, abutment, prosthetic screw, multi-unit, temporary — for Nobel Biocare, Straumann, Neodent, Zimmer Biomet, BioHorizons, Osstem, MIS, Megagen and Hiossen. Each value is quoted from the manufacturer’s document, and every gap is stated as a gap.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
FINAL ABUTMENTNobel Biocare35 Ncm
FINAL ABUTMENTStraumann35 Ncm
FINAL ABUTMENTNeodent20 Ncm
EDITORIAL VISUALIZATION

THE SYSTEMS

Same connection type, different numbers.

A conical connection at 20 Ncm and an internal hex at 35 Ncm can sit in the same arch. The brand page tells you which is which before the wrench is set.

SystemPlacementFinal abutmentProsthetic screwMulti-unit
ENVISTANobel BiocareConical Connection (CC) · Tri-channel (NobelReplace) · External hex (Brånemark)35–70NobelActive 3.5 · 4.3 · 5.0 · 5.5 — maximum 7035Final abutment screw, all platforms except 3.015Prosthetic screws for final restorations on Multi-unit35 · 15Multi-unit Abutment → implant: straight 35 · 17° and 30° angled 15See values
STRAUMANN GROUPStraumannCrossFit (BL/BLT) · TorcFit (BLX) · TLX NT/RT/WT≥ 35BLX · TLX, immediate function35Final abutment, all BL / BLX / TLX platforms, straight and angled15Occlusal screw on Screw-retained Abutment35Screw-retained Abutment, straight and angled (basal screw)See values
STRAUMANN GROUPNeodentGrand Morse (GM) — one prosthetic platform for Helix, Drive, Titamax, Helix Long and Zygoma32–60All GM implants — immediate loading ≥ 32; maximum 6020GM Exact · Titanium Base · Universal · Pro PEEK · CoCr abutments10Copings and definitive prostheses on Mini Conical / Micro32 · 20GM Mini Conical → implant: straight 32 · angled (17°/30°/45°/60°) 20See values
ZIMVIEZimmer BiometTapered Screw-Vent internal hex (TSV · TM · TSX) · 3i Certain internal · 3i External hex≤ 90TSX — instrument limits30TSV · TM · TSX, implant level, single unit20TSV Tapered Abutment coping screws · abutment-level Ti-Base30TSV Tapered Abutment → implant, straight and angledSee values
MIS IMPLANTS TECHNOLOGIES (DENTSPLY SIRONA)MISConical connection (C1 · V3) · Internal hex (Seven · M4) · Multi-Unit · CONNECT35–60C1 · V3 · Seven · M4 — maximum 45 for Ø3.30, 60 from Ø3.7530Titanium prosthetic screw — cementable, transgingival, Ti-Base, Ti blanks, CONNECT Ø4 / Ø5.725Titanium prosthetic screw for Multi-Unit (MU-S0220) · gold screw 2030 · 20Multi-Unit → implant: straight 30 · 17° and 30° angulated 20See values
OSSTEM IMPLANTOsstemTS internal conical hex — mini and regular platforms (TS II · TS III · TS IV)≤ 40TS II · TS III fixtures — HA-coated under 35; adverse effects above 5520 · 30Final abutment screw — mini 20 · regular 30 (Transfer, Angled, FreeForm ST, GoldCast, ZioCera, Rigid, Convertible)20Cylinder screw — Multi Combination Cylinder · Convertible cylinder30 · 20Straight Multi / Convertible abutment 30 · Multi Angled abutment mini 20, regular 30See values
BIOHORIZONS CAMLOGBioHorizonsInternal hex (Tapered Internal · Tapered Pro · Tapered Plus · 3.0 · Tissue Level) · Conical (Tapered Pro Conical)≤ 120Tapered Internal · Tapered Pro, mounted (3inOne) implants30Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)15Prosthetic screw on Multi-unit30Multi-unit abutment → implant, straight and 17° angledSee values
MEGAGEN IMPLANTMegagenAnyRidge internal conical · Blue Diamond conical · AnyOne External hex≤ 60AnyRidge — engine stops at 40, ratchet to the surgical plan, never above 6025AnyRidge abutment screw M1.8 / M2.0 (two-piece abutments) — general guidance 25–3515Cylinder screw M1.4 on Multi-unit35 · 25Multi-unit Abutment [AR] → implant: straight 35 · 17° and 30° angled 25See values
OSSTEM IMPLANT (US)HiossenET / EK internal conical hex — mini and regular platforms · SS (Octa) · EM≤ 40ET · EK drilling protocol — no numeric maximum published20 · 30Final abutment screw — mini 20 · regular 30 (Transfer, Angled, Rigid, Cast, CAD/CAM Ti-base)20Multi Combination Cylinder · Esthetic-low cylinders · Multi Temporary Cylinder · Multi Ti-Base30 · 20Multi Abutment straight 30 (mini + regular) · Multi Angled Abutment mini 20, regular 30See values

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

HOW TO READ THESE PAGES

Quoted, labelled, dated.

Manufacturer document

The value is quoted from an IFU, surgical or prosthetic manual, or torque guide, and that document is named beside the row.

Not published — confirm IFU

The documents we read do not give a number for that component. The row says so instead of borrowing a value from a neighbour.

Revision year

Each brand page lists the documents read and the newest revision year. Your printed IFU wins over this page.

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01Is there a single torque value for all dental implants?+

No. Final abutment screws on the nine systems here range from 20 Ncm (conical connections such as Neodent Grand Morse two-piece abutments, BioHorizons Tapered Pro Conical, 3i Certain) to 35 Ncm (Straumann Bone Level, Nobel Biocare standard abutments). Prosthetic screws on multi-unit abutments are typically 10–20 Ncm. Placement torque is stated as floors, ceilings or device limits depending on the manufacturer.

02Why does the same brand list 15 and 35 Ncm for a multi-unit?+

Because there are two screws. The multi-unit abutment tightens into the implant (often 30–35 Ncm, sometimes lower for angled versions), and the prosthetic screw tightens the bridge onto the abutment (often 15 Ncm or less). Each brand page separates the two rows.

03What if my system is not listed?+

Use the manufacturer’s IFU. In TrazaLab the torque field accepts the manufacturer values and a “per manufacturer” option, so the case record still carries what was applied.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.