Re-tightening, as the manufacturer states it
Hiossen states no re-tightening interval in the 2026 catalog or the ET III brochure we read.
IMPLANT TORQUE · HIOSSEN
One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Hiossen documentation, with the document named beside it.
Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
EDITORIAL VISUALIZATION“The recommended drill speed is 800–1,200 RPM, with an implant insertion torque of ≤ 40 Ncm.” A manufacturer maximum is not printed in the documents we read.
THE VALUES
Rows are grouped in treatment order. Where Hiossen publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.
| Component | Applies to | Torque | From the document | Source |
|---|---|---|---|---|
| PlacementInsertion torque of the fixture | ET · EK drilling protocol — no numeric maximum published Driver: Fixture mount · torque wrench | ≤ 40Ncm | “The recommended drill speed is 800–1,200 RPM, with an implant insertion torque of ≤ 40 Ncm.” A manufacturer maximum is not printed in the documents we read. | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
| HealingHealing abutment · cover screw | Healing abutment — hand tighten (ET III brochure: 5–8) Driver: 1.2 hex hand driver · 0.9 hex for OB2 / OB3 caps | hand-tight | 2026 catalog: “Healing Abutment — Hand tighten.” ET III brochure (2017): “Tightening torque: 5–8 Ncm.” OB healing caps: “Use 0.9 Hex Hand Driver … Recommended tightening torque: 5–8 Ncm · Disposable. Single use only.” | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
| AbutmentFinal abutment screw at implant level | Final abutment screw — mini 20 · regular 30 (Transfer, Angled, Rigid, Cast, CAD/CAM Ti-base) Driver: 1.2 hex driver | 20 · 30Ncm | “Recommended tightening torque: Mini: 20 Ncm — Regular: 30 Ncm.” “After connecting, make sure that it is tightened with a torque wrench.” Torque driver: “30 Ncm (excessive torque causes fracture)”. | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
| Multi-unitMulti-unit abutment into the implant | Multi Abutment straight 30 (mini + regular) · Multi Angled Abutment mini 20, regular 30 Driver: Multi Abutment Outer Driver Pin-Type (MAODP) · 1.2 hex (angled) | 30 · 20Ncm | “Torque using Multi Abutment Outer Driver Pin-Type (MAODP) … Recommended tightening torque: 30 Ncm (Mini + Regular).” “Multi Angled Abutment … 1.2 Hex Hand Driver · Recommended tightening torque: Mini: 20 Ncm — Regular: 30 Ncm.” | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
| Prosthetic screwScrew that holds the prosthesis on the abutment | Multi Combination Cylinder · Esthetic-low cylinders · Multi Temporary Cylinder · Multi Ti-Base Driver: 1.2 hex driver | 20Ncm | “Component for Multi Abutment … Torque using 1.2 Hex Driver · Recommended tightening torque: 20 Ncm · Packing unit: Cylinder + Ti Cylinder Screw.” | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
| TemporaryTemporary abutment or coping | Temporary Abutment — mini and regular Driver: 1.2 hex driver | 20Ncm | “Temporary Abutment … 1.2 Hex … tightening torque: 20 Ncm (mini / regular).” | Hiossen 2026 Product Catalog — US edition (PC26HPCLTR1.1, 05.2026)Manufacturer document |
Hiossen states no re-tightening interval in the 2026 catalog or the ET III brochure we read.
Newest revision read: 2026. Manufacturers revise these documents; the revision printed on your IFU wins.
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
WHY THE SCREW LOOSENS
Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

Preload
ONE SITE, FIVE MOMENTS
Site 36, one fixture, one crown. The same Hiossen values from the table, placed on the moment each one is read off the wrench.
EDITORIAL VISUALIZATIONET · EK drilling protocol — no numeric maximum published
Healing abutment — hand tighten (ET III brochure: 5–8)
Final abutment screw — mini 20 · regular 30 (Transfer, Angled, Rigid, Cast, CAD/CAM Ti-base)
Multi Combination Cylinder · Esthetic-low cylinders · Multi Temporary Cylinder · Multi Ti-Base
Final abutment screw — mini 20 · regular 30 (Transfer, Angled, Rigid, Cast, CAD/CAM Ti-base)
Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.
AFTER THE CLICK
In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.
See the Digital RxOTHER SYSTEMS
Each page is built the same way: quoted values, the document beside them, and the gaps stated as gaps.
Final abutment · Conical Connection (CC) · Tri-channel (NobelReplace) · External hex (Brånemark)
Final abutment · CrossFit (BL/BLT) · TorcFit (BLX) · TLX NT/RT/WT
Final abutment · Grand Morse (GM) — one prosthetic platform for Helix, Drive, Titamax, Helix Long and Zygoma
Final abutment · Tapered Screw-Vent internal hex (TSV · TM · TSX) · 3i Certain internal · 3i External hex
Final abutment · Conical connection (C1 · V3) · Internal hex (Seven · M4) · Multi-Unit · CONNECT
Final abutment · TS internal conical hex — mini and regular platforms (TS II · TS III · TS IV)
Final abutment · Internal hex (Tapered Internal · Tapered Pro · Tapered Plus · 3.0 · Tissue Level) · Conical (Tapered Pro Conical)
Final abutment · AnyRidge internal conical · Blue Diamond conical · AnyOne External hex
Final abutment · ET / EK internal conical hex — mini and regular platforms · SS (Octa) · EM
QUESTIONS
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
For the values we read, yes: mini 20 Ncm and regular 30 Ncm for the final abutment, 30 Ncm for the straight Multi Abutment, 20 Ncm for the cylinder screw, hand-tight healing components, and 40 Ncm or less at placement. Hiossen is Osstem’s US catalog. The document that ships with your component still decides.
20 Ncm with the 1.2 hex driver, for the Multi Combination Cylinder, the Esthetic-low gold and plastic cylinders, the Multi Temporary Cylinder and the Multi Ti-Base. The Multi Abutment itself goes on at 30 Ncm with the pin-type outer driver.
No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.
The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.
In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.
AFTER THE CLICK
TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.