IMPLANT TORQUE · STRAUMANN

Straumann torque values.

One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Straumann documentation, with the document named beside it.

Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
CONNECTIONSCrossFit (BL/BLT) · TorcFit (BLX) · TLX NT/RT/WT
LINES COVEREDBone Level · BLT · BLX · TLX · Screw-retained Abutment (Pro Arch)
TrazaLabTORQUE SCALE StraumannManufacturer document
EDITORIAL VISUALIZATION
≥ 35Ncm
PLACEMENTBLX · TLX, immediate function

Floor, not ceiling. “Excessive insertion torque must be avoided because this can lead to resorption of the bone.” No numeric maximum is published; the Loxim transfer piece breaks at 80 Ncm.

DRIVERRatchet + Torque Control Device
Straumann BLX Implant System — Basic Information (702115, © 2024)

THE VALUES

Every component, its number, its document.

Rows are grouped in treatment order. Where Straumann publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.

ComponentApplies toTorqueFrom the documentSource
PlacementInsertion torque of the fixtureBLX · TLX, immediate function
Driver: Ratchet + Torque Control Device
≥ 35NcmFloor, not ceiling. “Excessive insertion torque must be avoided because this can lead to resorption of the bone.” No numeric maximum is published; the Loxim transfer piece breaks at 80 Ncm.Straumann BLX Implant System — Basic Information (702115, © 2024)Manufacturer document
PlacementInsertion torque of the fixtureBLT (CrossFit)
Driver: Loxim + Ratchet
> 35 → checkNcmIf more than 35 Ncm is reached before the implant is fully seated, check the osteotomy to avoid bone over-compression.Straumann BLT Surgical Procedures (NAMLIT.1043, © 2023)Manufacturer document
HealingHealing abutment · cover screwHealing caps · closure screws
Driver: SCS Screwdriver
15NcmTorque tables list 15 Ncm; the step-by-step text says “hand-tighten the healing abutment”. Both are Straumann.Straumann Bone Level Prosthetic Procedures (702061, © 2026)Manufacturer document
AbutmentFinal abutment screw at implant levelFinal abutment, all BL / BLX / TLX platforms, straight and angled
Driver: SCS Screwdriver + Ratchet + Torque Control Device
35Ncm“Tighten the screw to 35 Ncm.” Not differentiated by platform (NC, RC, RB, WB, NT, RT, WT).Straumann Bone Level Prosthetic Procedures (702061, © 2026)Manufacturer document
Multi-unitMulti-unit abutment into the implantScrew-retained Abutment, straight and angled (basal screw)
Driver: SCS Screwdriver
35NcmModel abutments must also be torqued to 35 Ncm; a hand-tight abutment misrepresents the oral fit.Straumann Screw-retained Abutments (702880, © 2026)Manufacturer document
Prosthetic screwScrew that holds the prosthesis on the abutmentOcclusal screw on Screw-retained Abutment
Driver: SCS Screwdriver
15Ncm“Tighten the occlusal screw to 15 Ncm.” Same value for the temporary coping on the abutment.Straumann Screw-retained Abutments (702880, © 2026)Manufacturer document
TemporaryTemporary abutment or copingTemporary abutment, implant level
Driver: SCS Screwdriver
15–35Ncm“Tighten the screw between 15 Ncm and 35 Ncm.” Protective caps: hand-tight. Scanbody screw: hand-tight, maximum 15 Ncm.Straumann Bone Level Prosthetic Procedures (702061, © 2026)Manufacturer document

Re-tightening, as the manufacturer states it

Straumann states no re-tightening interval in the documents we read (BLX, TLX, BLT, Bone Level prosthetics, Screw-retained Abutments).

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

ONE SITE, FIVE MOMENTS

Where the numbers appear in the treatment.

Site 36, one fixture, one crown. The same Straumann values from the table, placed on the moment each one is read off the wrench.

EDITORIAL VISUALIZATION
  1. 01PLACEMENT≥ 35NcmFixture seated

    BLX · TLX, immediate function

  2. 02HEALING15NcmHealing abutment on

    Healing caps · closure screws

  3. 03ABUTMENT35NcmFinal abutment torqued

    Final abutment, all BL / BLX / TLX platforms, straight and angled

  4. 04PROSTHESIS15NcmCrown screw seated

    Occlusal screw on Screw-retained Abutment

  5. 05RE-TIGHTENING35NcmSame value, second pass

    Final abutment, all BL / BLX / TLX platforms, straight and angled

Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.

AFTER THE CLICK

The value you applied is the value the laboratory should read.

In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.

See the Digital Rx
TrazaLabTREATMENT DESIGN · TOOTH 36Digital Rx
TREATMENT DESIGN · TOOTH 36
IMPLANT SYSTEMStraumann
PLATFORMBone Level
CONNECTIONCrossFit (BL/BLT)
ABUTMENTFinal abutment
TORQUE APPLIED35 Ncm
DIGITAL RXTorque applied · 35 Ncm
LABORATORY · PER-TOOTH SHEET36 · implant crown · torque 35 Ncm

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01Is the Straumann abutment torque different for NC and RC platforms?+

No. The Bone Level prosthetic procedures give 35 Ncm for the final abutment screw on every platform, including angled abutments. Only the healing components (15 Ncm) and the occlusal screw on a Screw-retained Abutment (15 Ncm) differ.

02What is the maximum insertion torque for Straumann BLX?+

Straumann publishes a floor, not a ceiling: at least 35 Ncm for immediate function, with a warning that excessive insertion torque can lead to bone resorption. The Torque Control Device scale ends at 80 Ncm, which is also the Loxim breaking point. Any ‘70 Ncm maximum’ you find elsewhere is not in the BLX or TLX documentation we read.

03Do these Straumann values replace the IFU?+

No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.

04Should I re-tighten after ten minutes?+

The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.

05Where does the applied torque go after the appointment?+

In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.