IMPLANT TORQUE · BIOHORIZONS

BioHorizons torque values.

One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from BioHorizons documentation, with the document named beside it.

Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
CONNECTIONSInternal hex (Tapered Internal · Tapered Pro · Tapered Plus · 3.0 · Tissue Level) · Conical (Tapered Pro Conical)
LINES COVEREDTapered Internal · Tapered Pro · Tapered Plus · Tapered 3.0 · Tissue Level · Tapered Pro Conical · Multi-unit / TeethXpress
TrazaLabTORQUE SCALE BioHorizonsManufacturer document
EDITORIAL VISUALIZATION
≤ 120Ncm
PLACEMENTTapered Internal · Tapered Pro, mounted (3inOne) implants

“When placing mounted implants, do not exceed 120 Ncm of torque”; then remove the 3inOne abutment and continue with the implant-level driver. No recommended minimum is stated. TeethXpress: hand-tighten the multi-unit screw if implant torque was below 30 Ncm.

BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)

THE VALUES

Every component, its number, its document.

Rows are grouped in treatment order. Where BioHorizons publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.

ComponentApplies toTorqueFrom the documentSource
PlacementInsertion torque of the fixtureTapered Internal · Tapered Pro, mounted (3inOne) implants≤ 120Ncm“When placing mounted implants, do not exceed 120 Ncm of torque”; then remove the 3inOne abutment and continue with the implant-level driver. No recommended minimum is stated. TeethXpress: hand-tighten the multi-unit screw if implant torque was below 30 Ncm.BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document
PlacementInsertion torque of the fixtureTapered Pro Conical: narrow · regular connection≤ 117 · ≤ 136NcmMaximum 117 Ncm for narrow-connection and 136 Ncm for regular-connection conical implants. “Tightening to a torque value greater than recommended may compromise the mechanical integrity of the dental implant driver.”BioHorizons IFU Conical Dental Implants (L01087 Rev E, JAN 2026)Manufacturer document
HealingHealing abutment · cover screwHealing abutment, all lines
Driver: .050" (1.25 mm) hex
10–15Ncm“Hand-tighten (10–15 Ncm) utilizing .050" (1.25 mm) Hex Driver.”BioHorizons Tapered Pro Conical Catalog & Surgical Manual (L02091 REV E, 2025)Manufacturer document
AbutmentFinal abutment screw at implant levelInternal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)
Driver: .050" (1.25 mm) hex · 4 mm square torque wrench
30Ncm“Abutment screw (PXAS) … Final torque: 30 Ncm.”BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document
AbutmentFinal abutment screw at implant levelTapered Pro Conical — conical abutments
Driver: .050" (1.25 mm) hex
20Ncm“Tighten to 20 Ncm.” A different connection from the internal-hex Tapered Pro.BioHorizons Tapered Pro Conical Catalog & Surgical Manual (L02091 REV E, 2025)Manufacturer document
Multi-unitMulti-unit abutment into the implantMulti-unit abutment → implant, straight and 17° angled
Driver: Multi-unit Hex Adapter · .050" hex
30Ncm“Torque the abutment screw up to 30 Ncm … Do not exceed 30 Ncm.”BioHorizons TeethXpress Technique Manual (L01021 REV G, 2019)Manufacturer document
Prosthetic screwScrew that holds the prosthesis on the abutmentProsthetic screw on Multi-unit
Driver: .050" (1.25 mm) hex
15Ncm“Tighten the prosthetic screws to 15 Ncm … Do not exceed 15 Ncm when tightening prosthetic screws.”BioHorizons Multi-unit Abutment Technique (Document 50130)Manufacturer document
TemporaryTemporary abutment or copingTemporary abutment: internal hex 30 · conical 20
Driver: .050" (1.25 mm) hex
30 · 20NcmPEEK (≤ 30 days) and TiN (> 30 days) temporary abutments on internal hex: 30 Ncm with the PXAS screw. Conical temporary abutment: 20 Ncm. Multi-unit cover caps and verification copings: hand-tight.BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document

Re-tightening, as the manufacturer states it

BioHorizons’ TeethXpress workflow states: “If the screws have been removed twice, replace screws. Torque prosthetic screws to 15 Ncm.” No time-based re-tightening statement was found.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

ONE SITE, FIVE MOMENTS

Where the numbers appear in the treatment.

Site 36, one fixture, one crown. The same BioHorizons values from the table, placed on the moment each one is read off the wrench.

EDITORIAL VISUALIZATION
  1. 01PLACEMENT≤ 120NcmFixture seated

    Tapered Internal · Tapered Pro, mounted (3inOne) implants

  2. 02HEALING10–15NcmHealing abutment on

    Healing abutment, all lines

  3. 03ABUTMENT30NcmFinal abutment torqued

    Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)

  4. 04PROSTHESIS15NcmCrown screw seated

    Prosthetic screw on Multi-unit

  5. 05RE-TIGHTENING30NcmSame value, second pass

    Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)

Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.

AFTER THE CLICK

The value you applied is the value the laboratory should read.

In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.

See the Digital Rx
TrazaLabTREATMENT DESIGN · TOOTH 36Digital Rx
TREATMENT DESIGN · TOOTH 36
IMPLANT SYSTEMBioHorizons
PLATFORMTapered Internal
CONNECTIONInternal hex (Tapered Internal
ABUTMENTFinal abutment
TORQUE APPLIED30 Ncm
DIGITAL RXTorque applied · 30 Ncm
LABORATORY · PER-TOOTH SHEET36 · implant crown · torque 30 Ncm

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01Is BioHorizons Tapered Pro 30 or 20 Ncm?+

Both exist, for two different implants. Tapered Pro (internal hex) uses the PXAS abutment screw at 30 Ncm. Tapered Pro Conical is a separate conical connection whose abutments tighten to 20 Ncm. The case record should carry the full product name.

02What is the BioHorizons multi-unit prosthetic screw torque?+

15 Ncm, and the technique document adds “do not exceed 15 Ncm”. The multi-unit abutment itself goes to 30 Ncm, straight or angled, again with a ‘do not exceed 30 Ncm’.

03Do these BioHorizons values replace the IFU?+

No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.

04Should I re-tighten after ten minutes?+

The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.

05Where does the applied torque go after the appointment?+

In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.