Re-tightening, as the manufacturer states it
BioHorizons’ TeethXpress workflow states: “If the screws have been removed twice, replace screws. Torque prosthetic screws to 15 Ncm.” No time-based re-tightening statement was found.
IMPLANT TORQUE · BIOHORIZONS
One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from BioHorizons documentation, with the document named beside it.
Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
EDITORIAL VISUALIZATION“When placing mounted implants, do not exceed 120 Ncm of torque”; then remove the 3inOne abutment and continue with the implant-level driver. No recommended minimum is stated. TeethXpress: hand-tighten the multi-unit screw if implant torque was below 30 Ncm.
THE VALUES
Rows are grouped in treatment order. Where BioHorizons publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.
| Component | Applies to | Torque | From the document | Source |
|---|---|---|---|---|
| PlacementInsertion torque of the fixture | Tapered Internal · Tapered Pro, mounted (3inOne) implants | ≤ 120Ncm | “When placing mounted implants, do not exceed 120 Ncm of torque”; then remove the 3inOne abutment and continue with the implant-level driver. No recommended minimum is stated. TeethXpress: hand-tighten the multi-unit screw if implant torque was below 30 Ncm. | BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document |
| PlacementInsertion torque of the fixture | Tapered Pro Conical: narrow · regular connection | ≤ 117 · ≤ 136Ncm | Maximum 117 Ncm for narrow-connection and 136 Ncm for regular-connection conical implants. “Tightening to a torque value greater than recommended may compromise the mechanical integrity of the dental implant driver.” | BioHorizons IFU Conical Dental Implants (L01087 Rev E, JAN 2026)Manufacturer document |
| HealingHealing abutment · cover screw | Healing abutment, all lines Driver: .050" (1.25 mm) hex | 10–15Ncm | “Hand-tighten (10–15 Ncm) utilizing .050" (1.25 mm) Hex Driver.” | BioHorizons Tapered Pro Conical Catalog & Surgical Manual (L02091 REV E, 2025)Manufacturer document |
| AbutmentFinal abutment screw at implant level | Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level) Driver: .050" (1.25 mm) hex · 4 mm square torque wrench | 30Ncm | “Abutment screw (PXAS) … Final torque: 30 Ncm.” | BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document |
| AbutmentFinal abutment screw at implant level | Tapered Pro Conical — conical abutments Driver: .050" (1.25 mm) hex | 20Ncm | “Tighten to 20 Ncm.” A different connection from the internal-hex Tapered Pro. | BioHorizons Tapered Pro Conical Catalog & Surgical Manual (L02091 REV E, 2025)Manufacturer document |
| Multi-unitMulti-unit abutment into the implant | Multi-unit abutment → implant, straight and 17° angled Driver: Multi-unit Hex Adapter · .050" hex | 30Ncm | “Torque the abutment screw up to 30 Ncm … Do not exceed 30 Ncm.” | BioHorizons TeethXpress Technique Manual (L01021 REV G, 2019)Manufacturer document |
| Prosthetic screwScrew that holds the prosthesis on the abutment | Prosthetic screw on Multi-unit Driver: .050" (1.25 mm) hex | 15Ncm | “Tighten the prosthetic screws to 15 Ncm … Do not exceed 15 Ncm when tightening prosthetic screws.” | BioHorizons Multi-unit Abutment Technique (Document 50130)Manufacturer document |
| TemporaryTemporary abutment or coping | Temporary abutment: internal hex 30 · conical 20 Driver: .050" (1.25 mm) hex | 30 · 20Ncm | PEEK (≤ 30 days) and TiN (> 30 days) temporary abutments on internal hex: 30 Ncm with the PXAS screw. Conical temporary abutment: 20 Ncm. Multi-unit cover caps and verification copings: hand-tight. | BioHorizons Tapered Family Catalog & Surgical Manual (L02024 REV G, 2020)Manufacturer document |
BioHorizons’ TeethXpress workflow states: “If the screws have been removed twice, replace screws. Torque prosthetic screws to 15 Ncm.” No time-based re-tightening statement was found.
Newest revision read: 2026. Manufacturers revise these documents; the revision printed on your IFU wins.
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
WHY THE SCREW LOOSENS
Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

Preload
ONE SITE, FIVE MOMENTS
Site 36, one fixture, one crown. The same BioHorizons values from the table, placed on the moment each one is read off the wrench.
EDITORIAL VISUALIZATIONTapered Internal · Tapered Pro, mounted (3inOne) implants
Healing abutment, all lines
Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)
Prosthetic screw on Multi-unit
Internal hex — PXAS abutment screw (Tapered Internal · Pro · Plus · 3.0 · Tissue Level)
Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.
AFTER THE CLICK
In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.
See the Digital RxOTHER SYSTEMS
Each page is built the same way: quoted values, the document beside them, and the gaps stated as gaps.
Final abutment · Conical Connection (CC) · Tri-channel (NobelReplace) · External hex (Brånemark)
Final abutment · CrossFit (BL/BLT) · TorcFit (BLX) · TLX NT/RT/WT
Final abutment · Grand Morse (GM) — one prosthetic platform for Helix, Drive, Titamax, Helix Long and Zygoma
Final abutment · Tapered Screw-Vent internal hex (TSV · TM · TSX) · 3i Certain internal · 3i External hex
Final abutment · Conical connection (C1 · V3) · Internal hex (Seven · M4) · Multi-Unit · CONNECT
Final abutment · TS internal conical hex — mini and regular platforms (TS II · TS III · TS IV)
Final abutment · Internal hex (Tapered Internal · Tapered Pro · Tapered Plus · 3.0 · Tissue Level) · Conical (Tapered Pro Conical)
Final abutment · AnyRidge internal conical · Blue Diamond conical · AnyOne External hex
Final abutment · ET / EK internal conical hex — mini and regular platforms · SS (Octa) · EM
QUESTIONS
Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.
Both exist, for two different implants. Tapered Pro (internal hex) uses the PXAS abutment screw at 30 Ncm. Tapered Pro Conical is a separate conical connection whose abutments tighten to 20 Ncm. The case record should carry the full product name.
15 Ncm, and the technique document adds “do not exceed 15 Ncm”. The multi-unit abutment itself goes to 30 Ncm, straight or angled, again with a ‘do not exceed 30 Ncm’.
No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.
The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.
In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.
AFTER THE CLICK
TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.