IMPLANT TORQUE · NOBEL BIOCARE

Nobel Biocare torque values.

One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Nobel Biocare documentation, with the document named beside it.

Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
CONNECTIONSConical Connection (CC) · Tri-channel (NobelReplace) · External hex (Brånemark)
LINES COVEREDNobelActive · NobelActive 3.0 · NobelParallel CC · NobelReplace CC · NobelReplace Tapered · NobelSpeedy · Brånemark System · Multi-unit Abutment
TrazaLabTORQUE SCALE Nobel BiocareManufacturer document
EDITORIAL VISUALIZATION
35–70Ncm
PLACEMENTNobelActive 3.5 · 4.3 · 5.0 · 5.5 — maximum 70

“Never exceed insertion torque of … 70 Ncm for NobelActive 3.5, 4.3, 5.0 and 5.5.” Immediate function: final torque 35–70 Ncm.

DRIVERImplant Driver · Manual Torque Wrench Surgical
NobelActive IFU1001 (EN_US rev. 03)

THE VALUES

Every component, its number, its document.

Rows are grouped in treatment order. Where Nobel Biocare publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.

ComponentApplies toTorqueFrom the documentSource
PlacementInsertion torque of the fixtureNobelActive 3.5 · 4.3 · 5.0 · 5.5 — maximum 70
Driver: Implant Driver · Manual Torque Wrench Surgical
35–70Ncm“Never exceed insertion torque of … 70 Ncm for NobelActive 3.5, 4.3, 5.0 and 5.5.” Immediate function: final torque 35–70 Ncm.NobelActive IFU1001 (EN_US rev. 03)Manufacturer document
PlacementInsertion torque of the fixtureNobelActive 3.0 · NobelParallel CC · NobelReplace · NobelSpeedy · Brånemark — maximum 45
Driver: Implant Driver · Manual Torque Wrench Surgical
35–45Ncm“NobelActive 3.0 Max 45 Ncm … All other Nobel Biocare implants … Max 45 Ncm.” NobelParallel CC: “insertion torque between 35 and 45 Ncm … Never exceed insertion torque of 45 Ncm.” NobelSpeedy: 20–45 Ncm.Nobel Biocare Torque Guide — Implants (catalog p. 326)Manufacturer document
HealingHealing abutment · cover screwHealing abutment · healing cap · plastic temporary abutment and coping
Driver: Unigrip Screwdriver
hand-tight“By hand: Plastic Temporary Abutment, Plastic Temporary Coping, Healing Abutment, Healing Cap.”Nobel Biocare Prosthetic Torque Guide (catalog appendix p. 353)Manufacturer document
AbutmentFinal abutment screw at implant levelFinal abutment screw, all platforms except 3.0
Driver: Unigrip Screwdriver · Manual Torque Wrench Prosthetic
35Ncm“All other abutments for Nobel Biocare implant systems — 35 Ncm.” NobelParallel CC: “recommended tightening torque is 35 Ncm. Caution: never exceed the recommended maximum tightening torque for the abutment screw.”Nobel Biocare Prosthetic Torque Guide (catalog appendix p. 353)Manufacturer document
AbutmentFinal abutment screw at implant levelAbutments for NobelActive 3.0 and CC 3.0
Driver: Unigrip Screwdriver
≤ 15Ncm“Abutments for NobelActive 3.0 — Max 15 Ncm.” IFU1001: “maximum prosthetic abutment tightening torque is 15 Ncm … Overtightening of the abutment may lead to a screw fracture.”NobelActive IFU1001 (EN_US rev. 03)Manufacturer document
Multi-unitMulti-unit abutment into the implantMulti-unit Abutment → implant: straight 35 · 17° and 30° angled 15
Driver: Screwdriver Machine Multi-unit (straight) · Unigrip (angled)
35 · 15Ncm“Tighten straight Multi-unit Abutments to 35 Ncm using the Screwdriver Machine Multi-unit … 17° Multi-unit Abutments to 15 Ncm using the Unigrip Screwdriver.” NobelActive manual: “Be sure not to exceed 15 Ncm for Multi-unit Angulated Abutment screw.”All-on-4 Treatment Concept Procedures Manual (2016 copy)Manufacturer document
Prosthetic screwScrew that holds the prosthesis on the abutmentProsthetic screws for final restorations on Multi-unit
Driver: Unigrip Screwdriver · Manual Torque Wrench Prosthetic
15Ncm“Prosthetic Screws for final restorations — 15 Ncm.” All-on-4: “Tighten all screws to 15 Ncm using Unigrip Screwdriver Machine and Manual Torque Wrench Prosthetic.”Nobel Biocare Prosthetic Torque Guide (catalog appendix p. 353)Manufacturer document
TemporaryTemporary abutment or copingEngaging Temporary Abutment Ti / PEEK 35 · Temporary Coping, Snap Coping, Slim and 3.0 temporaries 15
Driver: Unigrip Screwdriver
35 · 15NcmIFU1093: “Tighten to recommended torque of 35 Ncm” (Temporary Abutment Engaging CC, PEEK); “Connect and tighten the temporary restoration to 15 Ncm” (copings); Slim and 3.0: 15 Ncm. Temporary abutments are single use.Nobel Biocare Temporary Abutments IFU1093 (EN_GB rev. 06)Manufacturer document

Re-tightening, as the manufacturer states it

Nobel Biocare states no re-tightening interval in the torque guides, IFU1001, IFU1093, the NobelParallel CC manual or the All-on-4 manual we read.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

ONE SITE, FIVE MOMENTS

Where the numbers appear in the treatment.

Site 36, one fixture, one crown. The same Nobel Biocare values from the table, placed on the moment each one is read off the wrench.

EDITORIAL VISUALIZATION
  1. 01PLACEMENT35–70NcmFixture seated

    NobelActive 3.5 · 4.3 · 5.0 · 5.5 — maximum 70

  2. 02HEALINGhand-tightHealing abutment on

    Healing abutment · healing cap · plastic temporary abutment and coping

  3. 03ABUTMENT35NcmFinal abutment torqued

    Final abutment screw, all platforms except 3.0

  4. 04PROSTHESIS15NcmCrown screw seated

    Prosthetic screws for final restorations on Multi-unit

  5. 05RE-TIGHTENING35NcmSame value, second pass

    Final abutment screw, all platforms except 3.0

Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.

AFTER THE CLICK

The value you applied is the value the laboratory should read.

In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.

See the Digital Rx
TrazaLabTREATMENT DESIGN · TOOTH 36Digital Rx
TREATMENT DESIGN · TOOTH 36
IMPLANT SYSTEMNobel Biocare
PLATFORMNobelActive
CONNECTIONConical Connection (CC)
ABUTMENTFinal abutment
TORQUE APPLIED35 Ncm
DIGITAL RXTorque applied · 35 Ncm
LABORATORY · PER-TOOTH SHEET36 · implant crown · torque 35 Ncm

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01What is the abutment torque for NobelActive 3.0?+

15 Ncm maximum, not 35. The Prosthetic Torque Guide lists “Abutments for NobelActive 3.0 — Max 15 Ncm”, and IFU1001 repeats that overtightening may fracture the screw. Every other Nobel Biocare platform tightens the final abutment to 35 Ncm.

02Is the Nobel Biocare multi-unit abutment 35 or 15 Ncm?+

Both, by angulation. The straight Multi-unit Abutment is tightened to 35 Ncm with the Screwdriver Machine Multi-unit; the 17° and 30° angled abutments to 15 Ncm with the Unigrip. The prosthetic screw that holds the bridge on top is 15 Ncm in every case.

03Do these Nobel Biocare values replace the IFU?+

No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.

04Should I re-tighten after ten minutes?+

The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.

05Where does the applied torque go after the appointment?+

In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.