IMPLANT TORQUE · OSSTEM

Osstem torque values.

One implant site. Five moments where a number decides whether the screw holds: placement, healing, abutment, prosthesis, re-tightening. Every value below is quoted from Osstem documentation, with the document named beside it.

Scale 0–70 Ncm. Select a component to move the pointer. Values beyond the scale sit at its end.
CONNECTIONSTS internal conical hex — mini and regular platforms (TS II · TS III · TS IV)
LINES COVEREDTS II · TS III · TS IV · Multi Abutment · Multi Angled Abutment · Convertible Abutment
TrazaLabTORQUE SCALE OsstemManufacturer document
EDITORIAL VISUALIZATION
≤ 40Ncm
PLACEMENTTS II · TS III fixtures — HA-coated under 35; adverse effects above 55

“Recommended insertion torque: ≤ 40 Ncm.” “Excessive torque (greater than 55 Ncm) in the fixture placement can have adverse effects such as partial fracture or necrosis of the bone.” HA-coated: “insertion torque of the implant should be less than 35 Ncm.”

DRIVERFixture mount · torque wrench
Osstem TS System User Manual / Catalog (Osstem UK)

THE VALUES

Every component, its number, its document.

Rows are grouped in treatment order. Where Osstem publishes a hand-tight instruction instead of a number, the row says so. Where nothing is published, the row says that too.

ComponentApplies toTorqueFrom the documentSource
PlacementInsertion torque of the fixtureTS II · TS III fixtures — HA-coated under 35; adverse effects above 55
Driver: Fixture mount · torque wrench
≤ 40Ncm“Recommended insertion torque: ≤ 40 Ncm.” “Excessive torque (greater than 55 Ncm) in the fixture placement can have adverse effects such as partial fracture or necrosis of the bone.” HA-coated: “insertion torque of the implant should be less than 35 Ncm.”Osstem TS System User Manual / Catalog (Osstem UK)Manufacturer document
HealingHealing abutment · cover screwCover screw · healing abutment · impression coping — hand tightening, less than 10
Driver: 1.2 hex hand driver
hand-tight“Cover screw, Healing abutment, Impression coping, Bite index — Hand tightening (less than 10 Ncm).”Osstem TS Implant System — Prosthetic Procedure (2013)Manufacturer document
AbutmentFinal abutment screw at implant levelFinal abutment screw — mini 20 · regular 30 (Transfer, Angled, FreeForm ST, GoldCast, ZioCera, Rigid, Convertible)
Driver: 1.2 hex driver
20 · 30Ncm“Tightening torque — Mini: 20 Ncm — Regular: 30 Ncm.” Catalog: “Recommended tightening torque: 20 Ncm (mini), 30 Ncm (regular).”Osstem TS Implant System — Prosthetic Procedure (2013)Manufacturer document
Multi-unitMulti-unit abutment into the implantStraight Multi / Convertible abutment 30 · Multi Angled abutment mini 20, regular 30
Driver: Multi Abutment Outer Driver (MAOD) · 1.2 hex (angled)
30 · 20Ncm“Multi Angled Abutment … Use a 1.2 hex driver, Recommended tightening torque: 20 Ncm (mini), 30 Ncm (regular).” 2013 procedure: “Convertible abutment — Abutment: 30 Ncm” with the outer driver; the current UK catalog names the outer driver without a number.Osstem TS System User Manual / Catalog (Osstem UK)Manufacturer document
Prosthetic screwScrew that holds the prosthesis on the abutmentCylinder screw — Multi Combination Cylinder · Convertible cylinder
Driver: 1.2 hex driver
20Ncm“Multi Combination Cylinder … Using 1.2 hex driver, Recommended tightening torque: 20 Ncm.” 2013: “Abutment: 30 Ncm — Cylinder Screw: 20 Ncm.”Osstem TS System User Manual / Catalog (Osstem UK)Manufacturer document
TemporaryTemporary abutment or copingQuick Temporary Abutmentnot publishedThe TS catalog page for the temporary abutment prints no torque line in the documents we read. Not published — confirm the IFU packed with the component.Not published — confirm IFU

Re-tightening, as the manufacturer states it

Osstem states no re-tightening interval in the TS catalog or the 2013 prosthetic procedure we read.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

WHY THE SCREW LOOSENS

You torqued it. Then the metal settled.

Torque creates preload — the clamping force that keeps abutment and implant pressed together. Microscopic peaks on the thread surfaces flatten under that load, and part of the preload is gone before the patient leaves the chair. The literature calls it the settling effect, or embedment relaxation.

PRELOAD
The wrench clicks at the manufacturer’s value.

Preload

EDITORIAL VISUALIZATION

ONE SITE, FIVE MOMENTS

Where the numbers appear in the treatment.

Site 36, one fixture, one crown. The same Osstem values from the table, placed on the moment each one is read off the wrench.

EDITORIAL VISUALIZATION
  1. 01PLACEMENT≤ 40NcmFixture seated

    TS II · TS III fixtures — HA-coated under 35; adverse effects above 55

  2. 02HEALINGhand-tightHealing abutment on

    Cover screw · healing abutment · impression coping — hand tightening, less than 10

  3. 03ABUTMENT20 · 30NcmFinal abutment torqued

    Final abutment screw — mini 20 · regular 30 (Transfer, Angled, FreeForm ST, GoldCast, ZioCera, Rigid, Convertible)

  4. 04PROSTHESIS20NcmCrown screw seated

    Cylinder screw — Multi Combination Cylinder · Convertible cylinder

  5. 05RE-TIGHTENING20 · 30NcmSame value, second pass

    Final abutment screw — mini 20 · regular 30 (Transfer, Angled, FreeForm ST, GoldCast, ZioCera, Rigid, Convertible)

Shown as the same value as the abutment row. Whether and when to re-tighten follows the manufacturer statement quoted in the values section.

AFTER THE CLICK

The value you applied is the value the laboratory should read.

In TrazaLab the implant line on a tooth site keeps system, platform, connection, abutment and applied torque together. It travels into the Digital Rx and onto the laboratory’s per-tooth sheet. An unconfirmed torque is not blank — the laboratory sees it as a gap.

See the Digital Rx
TrazaLabTREATMENT DESIGN · TOOTH 36Digital Rx
TREATMENT DESIGN · TOOTH 36
IMPLANT SYSTEMOsstem
PLATFORMTS II
CONNECTIONTS internal conical hex — mini and regular platforms (TS II
ABUTMENTFinal abutment
TORQUE APPLIED20 · 30 Ncm
DIGITAL RXTorque applied · 20 · 30 Ncm
LABORATORY · PER-TOOTH SHEET36 · implant crown · torque 20 · 30 Ncm

QUESTIONS

Asked before the wrench is set.

Values are quoted from manufacturer documentation and can change by component, platform and revision. Confirm the IFU for the component and lot in your hand before torquing.

01Is the Osstem abutment torque 20 or 30 Ncm?+

It depends on the platform, not the abutment. The mini platform tightens to 20 Ncm, the regular platform to 30 Ncm, and that rule holds for Transfer, angled, FreeForm ST, GoldCast, ZioCera, Rigid and Convertible abutments. The platform belongs next to the brand on the case record.

02What is the maximum insertion torque for an Osstem TS III?+

Osstem recommends 40 Ncm or less and states that excessive torque greater than 55 Ncm can cause partial fracture or necrosis of the bone. HA-coated fixtures are held under 35 Ncm. The documents do not print a single hard maximum between those two lines.

03Do these Osstem values replace the IFU?+

No. They are quoted from the manufacturer documents listed on this page, with the revision year noted. The IFU packed with your component and lot is the instruction; this page puts the values next to the treatment step and the case record.

04Should I re-tighten after ten minutes?+

The literature on the settling effect widely advocates a second application at the same value after about ten minutes, and in vitro work shows it partially restores preload. The evidence for a fixed interval is limited. The manufacturer statement for this system is quoted in the values section; where it says nothing, this page does not turn that silence into a rule.

05Where does the applied torque go after the appointment?+

In TrazaLab it is a field on the implant line of the tooth site, next to system, platform, connection and abutment. The Digital Rx carries it to the laboratory, whose per-tooth sheet prints the torque. If it was never confirmed, the packet shows “Not on Rx” instead of a number.

AFTER THE CLICK

Read the value here. Keep it in the case.

TrazaLab records the implant line with its torque on the tooth site, writes it into the Digital Rx, and shows the laboratory exactly what was applied — or exactly what was not confirmed.

Choose the right next step

Start with one case.

Create an account now or walk through a complete case before deciding. Both paths are available today.

Already using TrazaLab? Log in

Choose based on what you need

ACCOUNTStart nowCreate an account through the functioning TrazaLab signup journey.
TOURUnderstand before decidingFollow the complete self-guided case without registration or a sales conversation.
CASECompare it with your flowUse the tour as a guide and activate TrazaLab when the process fits your team.